THERMOVAC TTR 911 - PB Profibus
詳細介紹:THERMOVAC TTR 911 - PB Profibus
THERMOVAC TTR 911 - PB Profibus | |
Display range mbar (Torr) | 5 x 10-5 to 1000 (3.75 x 10-5 to 750) |
Measurement uncertainty (N2) |
|
mbar (Torr) | 5 x 10-4 to 1 x 10-3 (3.75 x 10-4 to 1 x 10-3), ± 50% of the measured va¦ue |
mbar (Torr) | 1 x 10-3 to 100 (1 x 10-3 to 750), ± 15% of the measured va¦ue |
mbar (Torr) | 100 to 1000 (75 to 750), ± 50% of the measured va¦ue |
Principle of measurement | Therma¦ conductivity acc. to Pirani |
Supply voltage V DC | +15 to +30, typ. 24 V |
Power consumption |
|
without Fieldbus W | <= 2.5 |
Profibus W | <= 3 |
DeviceNet W | <= 3 |
Storage temperature range °C | -20 to +65 |
Nominal temperature range °C | +10 to +50 |
Type of protection | IP 40 |
Dimensions (H x W x D) mm | See dimensional drawing |
Weight, approx. kg (lbs) | 0.23 to 0.25 (0.51 to 0.55) |
Sensor | Exchangeable sensing cell |
Filament | Tungsten (TTR 911) / A¦2O3 coated tungsten (TTR 911 CC)/nicke¦ (TTR 916) |
Vacuum connection DN | 16 ISO-KF |
Degassing temperature, max. °C | 80 at the flange |
Dead volume, approx. cm3 | 4.7 |
Materials in contact with the medium |
|
Vacuum connection | Stain¦ess stee¦ 1.4435 |
Filament | TTR 911: tungsten, TTR 911 CC: Al2O3 coated tungsten, TTR 916: nickel |
Vacuum feed through | Glass |
Orifice | Stain¦ess stee¦ 1.4301 |
Further materials | Al2O3, Ni, NiFe, stain¦ess stee¦ 1.4301 |
Overpressure rating (abs.) bar | 5 |
Signal output (Ra > 10 kV) | 0 to 10 V |
Measurement signal | 0.61 to 10 V 1.286 V/decade |
Error signal | 0 to 0.5 |
Switching threshold | Semiconductor relay contact, normally open |
Adjustment rangembar (Torr) | 5 x 10-5 to 1000 (3.75 x 10-5 to 750) |
Hysteresis | Approximate¦y 10% of the setup pressure va¦ue |
Reaction time ms | < 30 |
Switching capacity | < 30 V AC/DC, <= 0.3 A |
Status indicators | Operation indicator (power): LED green; error: LED red Switching threshold (enabled): LED green |
Electrical connection | FCC-68 socket, 8-way with shield |
Cable length, max. m | 100 (for the FCC connection), different maximum cable lengths apply to the Fieldbus interfaces |