CERAVAC CTR Transmitters 101 N
詳細介紹:CERAVAC CTR Transmitters 101 N
Technical Data | CERAVAC Transmitter | |
CTR 101 N (45 °C heated) | ||
Full scale (FS) / Measurement range |
| 0,1 Torr / 1 x 10-5 − 0,1 Torr |
1 Torr / 1 x 10-4 − 1 Torr | ||
10 Torr / 1 x 10-3 − 10 Torr | ||
− | ||
100 Torr / 0,01 − 100 Torr | ||
1000 Torr / 0,1 − 1000 Torr | ||
Measurement uncertainty |
| 0.12% ± temperature effect |
0.15% ± temperature effect (0.1 Torr) | ||
Sensor Measurement principle |
| INCONEL® membrane Capacitance diaphragm gauge |
Supply voltage | V DC | +14 to +30 |
Power consumption | W | ≤ 11 (at operating temperature ≤ 8) |
Electrical connection |
| 15-pol. Sub-D |
Analog output Measurement range | V | 0 to 10 |
Interface |
| RS 232 |
Setpoints |
| 2 |
Status indicators |
| LED |
Max.. cable lenght | m | 30 |
Max.. overrange pressure | bar (hPa) | 3.1 (3100) |
Operating temperature range | °C (°F) | +15 to +40 |
Storage temperature range | °C (°F) | -20 to +80 |
Max.. bakeout temperature | °C (°F) | Not bakeable |
Max.. rel.. humidity | % n..c.. | 25 to 95 |
Installation orientation |
| Any |
Wetted part material |
| INCONEL®, Stainless steel 316 |
Dead volume, approx.. | cm3 | 6.29 |
Gewicht | g (lbs) | 669 (0.15) |
Protection class | IP | 40 |
CE certification |
| EMC Directive 2014/30/EEC |
Controller type |
| GRAPHIX ONE / TWO / THREE |
Temperature effects Zero of FS
|
%/°C |
|
0,0025 (1000/100/10/1 Torr | ||
0,005 (0,1 Torr) | ||
0,02 (0,1 Torr) | ||
Span of reading
| %/°C
| 0.01 (1000/100/10/1 Torr) |
0.03 (0.1 Torr) | ||
Response time (10% to 90% FS) | ms | 40 / 80 (1 Torr) / 120 (0.1 Torr) |